Supplied with Conductivity Cell K = 1.0, Temperature Probe,
Electrode Stand with Clamp, Dust Cover, DC Adaptor, Operational
& Instruction Manual, Manufacturer Test Certificate and Warranty Card.
Technical Specification :
- Conductivity Range : 0 to 20.00, 200.0, 2000 µS/Cm; 2 to 20.00, 200.0 ms/cm
- Conductivity Resolution : 0.001,0.01,0.1 µS/cm & 0.01, 0.001 mS/cm
- TDS Range : 0 to 10.00, 100.0, 1000 ppm; 10.00, 100.0 ppt(1ppt=10,000ppm)
- TDS Resolution : 0.001, 0.01, 0.1 ppm & 0.001, 0.01ppt
- Range Selection : Auto / Manual Ranging
- Temperature Range : 0 to 130.0ºC
- Temperature Resolution : 0.1ºC
- Temp. Compensation : Automatic & Manual
- Temperature Coefficient : 0-10.000 % per ºC (Adjustable)
- Suitable Cell Constant : 0.1, 0.5, 1.0 K
- Adjustable Cell Constant : 0.09 to 1.1 K
- Cond. to TDS Conv. Fac.: 0.40000 to 1.00000 (Adjustable)
- Backlit LCD Display : 16 x 2 line Alphanumeric
- Printer Capability : Yes - Parallel Port
- Power Requirement : DC Adaptor
Salient Features :
- Auto Ranging
- Scrolling Display of Conductivity at 25ºC
- Conductivity at actual Temperature & TDS
- Auto / Manual Temperature Compensation
- Last Calibration Data storage with date and reading
- Current Analysis Data Printing Facility with Calibration data
- 10 analysis data storage with corresponding Calibration data etc.
Analysis Report :
- Company Name : ANALAB (VADODARA)
- Analysis Date & Time : 02-FEB-2010 – 19.00
- CONDUCTIVITY: 100.0 mSm at 025.0ºC
- CONDUCTIVITY TO TDS FACTOR: 0.50000
- TDS: 050.13 ppt
|
Calibration Report :
- Company Name : ANALAB (VADODARA)
- Analysis Date & Time : 02-FEB-2010 – 18.59
- CELL CONSTANT: 01.000
- CALIBRATION RANGE: 5: 20mS to 200mS
- STANDARD SOLUTION CONDUCTIVITY: 00100.00 mSm
- MEASURED CONDUCTIVITY: 00078.609 mSm
- CAL. FACTOR: 1.2721
|
|