Automatic Leak Test Apparatus
(Without Vacuum Desiccator)

Model : µLTCal50 | Make : ANALAB | HSN Code : 90318000

Supplied with Vacuum Pump (Inbuilt with System), Power Cord, Operational & Instruction Manual with SOP, Calibration Test Certificate (Without Traceability Document) and Warranty Card.

Technical Specifications

  • Vacuum Controller : Microcontroller Based
  • Maximum Vacuum Level : Greater than 650 mmHg
  • Vacuum Stability :
    ● Set Value Below 400 mmHg : ±10 mmHg within 30 seconds
    ● Set Value Above 400 mmHg : ±5 mmHg within 30 seconds
  • Vacuum Hold Time : Variable from 1 sec. to 99 min
  • Vacuum Release Time : Variable from 1 sec. to 5 min
  • Vacuum Release Time : Variable from 1 sec. to 5 min
  • Vacuum Initial Delay Time : Variable from 1 to 60 second to achieve 100 mmHg
    (Run will be aborted if 100 mmHg is not achieved in 60 seconds)
  • Vacuum Release : Auto - after completion of Vacuum Hold Time
  • Vacuum Measurement : Digital
  • Vacuum Measurement Units : mmHg, bar, kPa, inHg, atm, psi, cmH2O, kg/cm2
  • Number of Sample : 1 to 99
  • Calibration : With NABL Certified Standard Digital Vacuum Gauge
  • Method : 40 method with View, Print & Delete facility
  • Pharmacopeia Complies : As per USP, IP, Ph. Eur.
  • Material of Construction : Stainless Steel
  • Display : 20 x 4 Line Alphanumeric Backlit LCD
  • Keypad : Membrane waterproof polycarbonate soft touch keypad compliance to UL/ROHS
  • Calibration Data Storage : Non-Volatile Memory which can store, print & view calibration date & time, company name, Username, set and actual reading of Vacuum, calibration done by, instrument serial number, instrument ID, model number etc.
  • Analysis Data Storage : More than 750 analysis data can be stored with company name, analysis date & time, record number, batch number / ID, sample name, username, assigned method number, numbers of samples, set vacuum, vacuum hold time, vacuum release time, vacuum calibration date, Instrument model number, serial number, instrument ID etc. Secondly, analysis data can print directly from analysis mode without saving analysis data in instrument.
  • Searching of data : With record number, batch number, date, date range and by scrolling.
  • Report Formation : As per GLP Compliance
  • Password Protection : Date & Time, Calibration Mode, Delete Stored Analysis Data and Delete Method.
  • Printer Interface : Centronics Parallel Port / Serial (RS-232 C) Port
  • Printer : 40 / 80 Column Dot Matrix Printer (Printer to be procure optionally)
  • Power Supply : 230VAC ± 10%, 50Hz