Tablet Hardness Tester (4IN1)

Model : µTHCal50 | Make : ANALAB | HSN Code : 90248091

Supplied with Power Cord, Operational & Instruction Manual with SOP, Calibration Test Certificate (Without Traceability Document) and Warranty Card.

Technical Specifications

  • Display of Results : Thickness, Diameter, Hardness and weightData
  • Testing range thickness and Diameter : 0.5 - 35.00 mm in samemechanism with measuring units either in mm/inch (No need For extension module for thickness measurement) with accuracy ofbetter ± 0.05 mm
  • Hardness : 5.0 - approx. 500 N (Newton) with accuracy of 1 N andalso 800N option
  • Force Rate : Adjustable for linear force / linear speed increase
  • Sample Throughput : Throughput is approximately 6-10 samplesper minute.
  • Number of tests for statistics : Up to 100
  • Report Formation : As per GLP Compliance
  • Method/Protocols : 10 different drug protocols can be saved.
  • Cal. Procedure : Needs 10 mm reference block and 10 kg reference weight (certified)
  • Display : Graphic with Pixels of 240 X 128.
  • Keyboard : Membrane waterproof polycarbonate soft touch keypad compliance to UL/ROHS
  • Pharmacopeia Complies : As per USP-1217 and EP-2.9.8
  • Hardness Range : 5-500N (800N Option Available)
  • Hardness Accuracy : >1N
  • Diameter and Thickness Range : 0.5 - 35 mm
  • Diameter Accuracy : >± 0.05 mm
  • Measuring Units : Thickness and Diameter in either mm or inches
  • Measurable Units : N (Newton) KP (Kilo Pond, 1KP = 9.81N) SC (Strong Cobb 1KP = 1.43Sc)
  • Force Rate : Constant (Linear) Force or Constant (Linear) Speed
  • Number of Test for Statistics : Up to 100
  • Stored Methods : Up to 10
  • Calibration Procedure : Requires 10mm Reference block and 10Kg Reference Weight (certified)
  • Load cell Linearity : Calibration with certified weight 5, 10, 15 kg
  • Distance Validation : 10mm Cert. Block (for thickness anddiameter) RS-232 interface with external Balance connectivity, Mettler, sartorius or Shimadzu analytical balance.
  • “Adjustment Free” Testing Jaws : Requires no adjustments ofthe test jaws that are designed to suit all tablet shapes and Sizes with a unique grooved design for proper alignment of each tabletbefore testing.
  • Over Range Alert : "Alert" will display if the tablet or sample isabove the defined hardness range with the option to continue or abort test.
  • Calibration Technique : No Tilt (User Friendly)
  • Distance Validation : 10 mm Certified Block (For Thickness andDiameter)
  • Load cell Linearity : 3 point Linear Calibration with certified weight (5, 10, 15 Kg)
  • Load cell Diagnostic Check : Yes (Check the Load Cell signal falls within range before making any further Calibration)
  • Self-Test : Yes
  • Fault Localization : Yes (Text Format)
  • Load cell Over load Protection : Yes
  • Password Protection : Yes (For Advance Settings)
  • Factory Setting Reset : Yes
  • Real Time Clock : Yes
  • Number of Tablet Detection : Yes
  • Tracking Deletion : Yes
  • On Screen Statistic View : Yes
  • Printer Interface : Serial & Centronics Parallel Port – Dot Matrix
  • Interface Port : USB Type A (2), USB Type B,LAN
  • Serial Interface : For Balance Connection of Mettler, Sartorius ORShimadzu Analytical Balance
  • Printer Selection : 40 / 80 column (Printer have to procure optionally)
  • Dimensions : Approx. 360mm x 570mm x 370mm (D x W x H)
  • Packaging Dimensions : Approx. 480mm x 680mm x 500mm (D xW x H)
  • Net Weight : Approx. 8kg
  • Gross Weight : Approx. 13kg
  • Power Supply : 230VAC ± 10%, 50/60 Hz